Main Function:
Sand dust test chamber is used for testing the sealing performance of case, free settlement of dust, ability to work in an environment without obvious air flow, and in outdoor places with strong airflow and large amount of dust and similar places for electricalengineering, electronics, automotive, motorcycle and other products.
Parameter:
Model |
DT-1000 |
|
Sample Limitation |
This test equipment prohibits from testing samples of flammable, explosive and volatile substances; and storage of corrosive substances, organisms, and strong electromagnetic emission sources. |
|
Volume, Weight, Size |
||
Volume |
1000L |
|
interior size |
W1000×H1000×D1000mm |
|
Weight |
About 350kg |
|
Working Noise |
≤75db (Measured in 1.5-meters front of the machine and 1.5 meters height from the ground) |
|
Air Switch |
63A |
|
Working Voltage |
AC220V50/60HZ1 phase 3-wire 6 square copper core wire |
|
Sample Power Supply (Optional) |
1、DC12V24V36V20A power supply 2、Reserve electrifiable interface terminals, configured by customers. 3、Dust-proof socket is installed in the chamber. The power supply is AC220V20A, which is controlled by programmable timer. |
|
Sample Vacuum Device (optional) |
The air pressure in the test chamber is lower than the outside, and a vacuum pump device is required for the test sample (IP5/IP6 test) The air pressure in the chamber is the same as the outside. No vacuum pump device is required. |
|
Power Supply and Conditions |
AC220V single phase; voltage allowable fluctuation range is 5%; frequency allowable fluctuation range is 50/60HZ; protective grounding resistance is less than 4Ω, requiring the user to equip the device with the corresponding capacity of air or fuse-free switch at the installation site, and this switch must be independently controlled, when the equipment is used to place the power sample in the box. Please use the external power supply for the sample power supply, and do not use the power directly. |
|
Performance Index: |
||
Environmental Conditions |
The environment temperature is 5°C ~ 30°C, relative humidity ≤ 85% (no condensation), the value measured under the conditions in the test chamber without sample. |
|
Temperature and Humidity Range |
Temperature range: normal temperature RT + 15 ° C (60 ° C is optional); humidity range: laboratory humidity (normal humidity; only display, cannot be controlled) |
|
Control Accuracy |
Temperature resolution:0.1℃ Uniformity of temperature distribution: ±2℃(Optional) Temperature Fluctuation: +0.5 (Optional) |
|
Specification parameter |
Mesh: Standard 75um aperture, if the customer's special requirements can be matched *large aperture 200um Air velocity: ≥7.5m/s Dust-blowing concentration:2kg~4kg/m3(GB/IP5X~6X) The standard machine uses talcum powder. If the customer needs to use other dust, the company will not provide it at the first time. |
|
Execution Criteria and Test Conditions |
GB2423.37-2006 IEC60068-2-68-1994 GB4208-2008/IEC60529:2001 level case protection GJB150.12-2009 Sand dust test method (dust test) GB7000-1-2007 Chapter 9 Provisions - Case protection test GB10485 Dust-proof Test Method and Equivalent JIS, DIN, MIL Test Standards |
|
Structure |
Interior Materials: 304# Stainless steel 1.0mm thick External Materials: Cold rolled steel plate 1.0mm thick paint treatment or 304 stainless steel The funnel-shaped inner case of the chamber. Dust-shaking motor assists dust recovery. Control chamber and test chamber are integral style. Special dust collector |
|
Door |
Single door, explosion-proof handle Atomized transparent viewing window Lights of viewing window Viewing window wiper |
|
Viewing Window |
On the door, size 250X350mm |
|
Control Panel |
Temperature controller, operating indicator, switch |
|
Machine Room |
Dust blower, dust blowing outlet |
|
Sample Rack |
304# stainless steel punching into a mesh, easy to use, material rack spacing can be adjusted |
|
Cabinet for Power Distribution |
General power circuit breaker, controller, distribution board, radiator, overtemperature protector, dust-shaking motor |
|
Heating System |
Temperature balance is achieved by controlling the heating with a scaly heat-dissipating heating tube P.I. D |
|
Dust blowing mode |
Special high-speed blower blows test dust into the test space |
|
Shaking dust mode |
Vibrating motor recovers the dust attached to the inner wall of the machine to the dust collector. |
|
Control System |
||
Controller |
Imported LCD display touch screen temperature controller Control temperature, blowing time, floating time and shaking time respectively separately Static time and cycle period, etc. |
|
Temperature Input Mode |
Human-machine interface, touch input |
|
Program Capacity and Control Function |
10 different groups of programs that can be used. Each group of programs can compile 10 conditional segments with a total of 100 segments. Set 99H59M every time, you can repeat the cycle command 50 times. Single setting condition time 999H59M With power-off program, memory starts automatically after power-on, and RS-485 communication interface The program is made in a dialogue style with editing, clearing, inserting, etc. Date, time adjustment function, screen locking function. Automatic adjustment of freezing capacity Reservation startup and shutdown function Display curves during program execution |
|
Communication Interface |
RS-232 communication interface can be used as monitoring and remote-control system and record test data. |
|
Operation Mode |
Program/Valuating |
|
Setting Mode |
Chinese/English interface, touch input |
|
Setting Range |
*The limits of the large temperature range are 5°C |
|
Power Failure Memory |
The power failure recovery mode can be set as: hot start/cold start/stop. |
|
Scheduled boot |
Boot-up time can be set at will |
|
Temperature Measurement |
PT100 Platinum resistance |
|
Curve Recording Function |
RAM with battery protection, which can save setting value, sampling value and sampling time of equipment; * Large recording time is 60 days. |
|
Software Usage Environment |
IBMPC Compatible Computer, CPU above P II, memory above 128M, Simplified Chinese Win or Simplified Chinese Windows XP Operating System |
|
Temperature control mode |
Anti-integral saturation PID, BTC balance thermostat control method (temperature test equipment) |
|
Protection Device |
||
Chamber |
Extreme over temperature, fan, motor overheating relay |
|
Power Supply |
Leakage protection, overload and short circuit protection |
|
Storage environment requirements |
Equipment environment temperature should be kept within 0 °C ~ +30 °C |
|
Installation site requirements |
The left and right side wall panels of the chamber shall not be less than 800mm from the wall, and the distance from the front wall to the wall shall not be less than 1500mm. When the equipment is transported, enter the doorway or passage, elevator, etc., whether it can enter, so as not to affect the progress of your project. |
|
transportation mode |
The whole machine transportation |
|