中文Language Search
Home > Products > Environmental Simulation Units > Climate Simulation Test Equipment > Fast-alternating high/low temp-damp-heat test chamber
Environmental Simulation Units
< Back

Fast-alternating high/low temperature-damp-heat test chamber

This machine specializes in testing the heat, cold, dry and moisture resistance of various materials. It can choose Chinese or English LCD to display touch screen, which is easy to operate and edit program. It can display complete system operation data, execution and program curve. When malfunctions occur in operation, the screen will automatically display the causes of the failure and provide troubleshooting methods. 

1.  Product Overview: 

This equipment is based on MIL-STD-2165 and IEC 600868-2-14. When performing temperature cycling stress screening test, it should be started from normal temperature, which is different from the traditional thermal shock test and temperature cycle test. The products can be tested are below: 

·      Material flaws: coating, material, wire crack cracking particles, chemical contamination, seal failure, etc.

·      Electronic flaws: increased contact resistance of solder joints, short circuit caused by solder joints, poor short circuit of capacitor transistor, IC damage, rupture of resistor, rupture of mica insulation gasket, drift of component parameters, etc.

·      Production flaws: joint or screw slack, poor solder joint contact, improper assembly, short circuit, improper component installation, misuse of components, etc.

 

2.  Specification: 

Model

GROQ165

GROQ408

GROQ800

GROQ1000

Interior Size (cm) W*D*H

50×50×660

60×80×100

100×80×100

100×100×100

External Size (cm) W*D*H

110×120×190

120×150×220

160×150×220

160×180×220

Parameter

Temperature Range

A:-20150 B:-40150 C:-70150

Humidity Range

2098R.H

Fluctuation/Uniformity

±0.5/±2

Humidity Deviation

2-3R.H

Temperature Rise Speed

0.115/min

Temperature Fall Speed

0.115/min

Control System

Controller

GRO1000 temperature instrument, 7-inch color LCD touch screen

Range of Accuracy

Setting accuracy: temperature 0.1 C, humidity 1% R. H. 

Indicating accuracy: temperature 0.1 C, humidity 1% R. H.

Temperature and Humidity Sensor

Platinum resistance PT100

Heating System

Fully independent system, nickel-chromium alloy electric heater

Humidification System

External isolated all stainless-steel shallow surface evaporative humidifier

Dehumidification System

Laminar contact dehumidification at dew point temperature of evaporator coil

Water Supply System

Humidification water supply can be automatically controlled and recover residual water, so as to save water and reduce consumption.

Refrigeration System

 French-made "Tecumseh" fully-enclosed air-cooled single-stage/cascade compressor

Circulatory System

Temperature-resistant and low noise air-conditioning motor, multi-blade centrifugal wind turbine

Materials

Interior Materials

High quality carbon steel plate, phosphating electrostatic spraying/

External Materials

Stainless-steel high-quality mirror panel

Thermal Insulation Material

Polyurethane rigid foam/superfine fiberglass sponge

Insulated Door Frame

Double-layer high and low temperature resistant silicone rubber door seal

Standard Equipment

Multi-layer heating defrosts with illuminated glass window 1 set, sample rack 2 pcs, test lead hole (25, 50, 100mm) 1pc

Protections

Leakage, short circuit, over temperature, water shortage, motor overheating, compressor overpressure, overload, overcurrent protection / controller power failure memory

Power Supply Voltage

AC380V±10 50±0.5Hz  three-phase five-wire system

Temperature/humidity of environment while using

5+30 ≤85%R.H

Note:

1. The above data are measured at environment temperature (QT) 25 ° C under no load conditions in the studio.

2, according to the user's specific requirements can be customized non-standard high and low temperature alternating heat and humidity test room

This technical information is subject to change without notice.


Download PDF