Applications:

This product is ideal for aerospace products, information electronic instruments, materials, electricians, electronic products, various electronic components in the high and low temperature or humid and hot environment to test their performance indicators.

Chamber Structure

The chamber body is CNC machine tool with non-reaction handle, which is easy to operate.

The interior of the chamber is imported high-grade stainless steel (SUS304) mirror panel, and the external is coating with A3 steel plate, which improves the appearance and cleanliness.
The water supply tank is placed in the lower right part of the chamber and has automatic protection of water shortage, which makes it easier for operators to replenish water sources.
The large viewing window is equipped with an illuminating light to keep the inside of the chamber bright, and the inside is clearly to observe at any time by using the tempered glass embedded in the heating body.

The humidifying control system pipeline is separated from the control circuit board in order to avoid malfunction for water leakage in the humidifying pipeline and improve safety.
The waterway system pipeline circuit system is convenient for maintenance and overhaul.
Double-layer high-temperature resistant high-strength sealing strip between the door and the chamber to ensure the sealing of the test area. 
The left side of the chamber is equipped with a test hole with a diameter of 50mm or 100mm, which can be used for external test power wires or signal wires.
High-quality fixed PU movable wheel at the chamber bottom.

Cooling System

The refrigerating machine of chamber is French-made "Tecumseh" fully-enclosed compressor.
Freezing system with unit or binary low temperature loop system design. 
The multi-wing blower can strongly circulate the air to any angle, and the temperature distribution in the testing area can be evenly distributed;
The design of air circulation, the air pressure and the air speed circulating outlet meet the test standard, and also can make the temperature stabilization time fast by instantaneously opening the door;
Temperature rising, cooling and system independence can improve efficiency, reduce test cost, increase life and reduce failure rate;

Control System

The test chamber temperature and humidity control instrument adopts self-developed temperature meter, 7’’ HD true color LCD touch screen, simple screen operation, easy program editing, no input key, direct touch screen option.
The controller operation interface is available in both English and Chinese, and the real-time operation graph can be displayed on the screen. 

With a capacity of 100 sets of 1000 segments and 999 cycles, the maximum value is set to 99 hours and 59 minutes per time. 
After the data and test conditions are entered, the controller has a screen lock function to avoid shutdown by manual touch.
With RS-232 or RS-485 communication interface, you can design programs on the computer, monitor the test process and perform functions such as automatic power on/off. With automatic calculation function, the temperature and humidity change conditions can be corrected immediately, making the temperature and humidity control more precise and stable.

Standard

GB/T2423.1-2008    GB/T2423.2-2008    GB/T2423.3-2006    GB/T2423.4-2008

Specifications:

Model

GRO165

GRO408

GRO800

GRO1000


Interior Size (cm)

W*D*H

50×50×660

60×80×100

100×80×100

100×100×100


External Size (cm) W*D*H

110×120×190

120×150×220

160×150×220

160×180×220


Maximum Power (Binary model)

5(KW)

6.0(KW)

9.0(KW)

10.5(KW)


Parameter

Temperature Range

A:-20150 B:-40150 C:-70150


Humidity Range

2098R.H


Fluctuation/Uniformity

±0.5/±2


Humidity Deviation

2-3R.H


Temperature Rise Speed

0.13.0/min


Temperature Fall Speed

0.13.0/min

Control System

Controller

GRO1000 thermometers, 7’’ HD true color LCD touch screen

Range of Accuracy

Setting accuracy: temperature 0.1 C, humidity 1% R. H. 

Indicating accuracy: temperature 0.1 C, humidity 1% R. H.

Temperature and Humidity Sensor

Platinum resistor PT100

Heating System

Fully independent system, nickel-chromium alloy electric heater

Humidification System

External isolated all stainless-steel shallow surface evaporative humidifier

Dehumidification System

Laminar contact dehumidification at dew point temperature of evaporator coil

Water Supply System

Humidification water supply is automatically controlled, and can recover residual water, save water and reduce consumption. 

Refrigeration System

French-made "Tecumseh" fully-enclosed air-cooled single-stage/multi-stage compressor refrigeration

Circulatory System

Temperature-resistant and low noise air-conditioned motor. Multi-blade centrifugal wind turbine

Materials

Interior Materials

High quality carbon steel plate. Phosphating electrostatic spray treatment / SUS304 stainless steel matte line hairline treatment

External Materials

SUS304 stainless steel mirror panel

Thermal Insulation Material

Polyurethane rigid foam / superfine glass fiber cotton

Insulated Door Frame

Double-layer high and low temperature resistant silicone rubber door seal

Standard Equipment

Multi-layer heating defrosts with 1 set of illumination glass window, 2 sample racks, 1 test lead hole (25, 50, 100mm)

Protection

Leakage, Short Circuit, Overtemperature, Water Shortage, Motor Overheating, Compressor Overpressure, Overload, Overcurrent Protection/Controller Outage Memory

Power Supply Voltage

AC380V±10 50±0.5Hz three-phase five-wire system

Temperature/humidity of environment while using

5+30 ≤85R.H

Note:

1. The above data were measured at ambient temperature (QT) 25 C. without load in the chamber.

2. Non-standard high and low temperature alternating humidity and heat laboratory can be customized according to the specific requirements of users.

This technical information is subject to change without notice.